Aschroft
Ashcroft (877)-274-2763
toll free - 24 Hours a Day
ashcroft home products about us contact us links safety



 
Ashcroft   Type RXLdp Ultra-Low
Differential
Transducer / Transmitter
  • Certified 1% accuracy
  • High overpressure protection
  • Stainless steel & Lexan* NEMA 1 construction
  • 4-20mA output signals plus an assortment of voltage outputs
  • Easy installation
      - Small size
      - Conduit compatible
      - Fits standard electrical
         Workboxes

 

  The Ashcroft® RXLdp introduces a variable capacitance sensor within a glassclad silicon chip. The patented Si-Glas™ technology combines the inherent high sensitivity of a variable capacitance transducer with the repeatability of a micro-machined, ultra-thin silicon diaphragm

The Ashcroft Si-Glas sensor now enables precise measurement and control of very low pressure. Although the ultra-thin silicon diaphragm deflects only a micron, the sensor is 100 times more sensitive to pressure than available silicon piezoresistive pressure sensors.

The Si-Glas sensor is composed of only sputtered metals and glass molecularly bonded to silicon. There are no epoxies or other organics in the sensor to contribute to drift or mechanical degradation over time.
RXLdp Transducer

PERFORMANCE CHARACTERISTICS
Standard Ranges (Inches W.C.)
Unidirectional Ranges:
Differential or Gage
0/0.1 0/1.0 0/3.0 0/25.0
0/0.25 0/1.5 0/5.0 0/50.0
0/0.5 0/2.0 0/10.0  
0/0.75 0/2.5 0/15.0  
Bidirectional Ranges:
Compound
0/0.5 0/1.0 0/5.0  
0/0.1 0/2.0 0/10.0  
0/0.25 0/2.5 0/25.0  
0/0.5 0/3.0    
Consult factory for nonstandard ranges.
Accuracy Class (F.S.):    1%
Nonlinearity  
  Terminal Point* ±0.8
  Best fit straight line(BFSL) ±0.6
Hysteresis ±0.5
Nonrepeatability ±0.10
* Including hysteresis
Stability: ±0.5%F.S./yr
ENVIRONMENTAL CHARACTERISTICS
Temperature Limits:
Storage: -40 to +180°F
Operating:   0 to +160°F
(10-95% R.H. noncondensing)
Compensated Range:  40 to +160°F
Thermal Coefficients:
ZERO ±0.025%F.S./°F
SPAN ±0.025%F.S./°F
 
FUNCTIONAL CHARACTERISTICS
Overpressure Limits (F.S.):
Proof 15 psig
Burst 50 psig
Mounting Position Error:
0.5"W.C. and higher =<0.1%F.S.
Below 0.5"W.C. =<0.25F.S.
ELECTRICAL SPECIFICATIONS
Output Signal: Power:
4-20mA (2-wire) 12-36 Vdc
1-5 Vdc 12-36 Vdc
1-6 Vdc 12-36 Vdc
0-5 Vdc 12-36 Vdc
0-10 Vdc 12-36 Vdc
Output Signal is independent of power supply changes.
Response Time:
250msec (factory set)
Warm-up time:
less than one second
Reverse Wiring Protected
12-36 Vdc range without effect on the output signal
Supply Current:
<6mA vor voltage output
 
PHYSICAL CHARACTERISTICS
Weight:
4.5 oz. NEMA 1 Case
MATERIALS:
Case:
Series stainless steel / Lexan
Pressure Connections:
dual barbed, 1/4 stainless steel
Media:
Clean, dry and noncorrosive gas (consult factory for use on other media)
NOT FOR USE ON LIQUIDS

OPTIONS
Back plate adapter (XRK)
Calibration report (XRH)
Nonstandard calibration (XCL)
NOTES:
Consult factory on other pressure range, temperature copensation, packaging variations or response time available

PART NUMBER FOR A TRANSDUCER/TRANSMITTER:
Select: RX 7 MB2 05 ST P1IW XRK
  Type
(rXLdp)
Accuracy
(7) 1.0%, ±0.025%/°F
Pressure Connection
(MB2) ¼ Barbed
(MB1) No Case Option
Output Signal
(05) 0/5Vdc
(10) 0/10Vdc
(15) 1/5Vdc
(16) 1/6Vdc
(42) 4-20mA
Electrical Termination
(ST) Screw Termination
Pressure Range
Diff. or Gauge:
(P1IW) 0.10"WC
(P25IW) 0.25"WC
(P5IW) 0.50"WC
(P75IW) 0.75"'WC
(1IW) 1.00"WC
(1P5IW) 1.5"WC
(2IW) 2.00"WC
(2P5IW) 2.50"WC
(3IW) 3.00"WC
(5IW) 5.00"WC
(10IW)10.00"WC
(25IW) 25.00"WC
(50IW) 50.00"WC


Compound:
(P1IWL) ±0.10"WC
(P25IWL) ±0.25"WC
(P5IWL) ±0.50"WC
(1IWL) ±1.00"WC
(2P5IWL) ±2.50"
(5IWL) ±5.00"WC
(10IWL) ±10.00"WC
(25IWL) ±25.00"WC
(50IWL) ±50.00"WC
Options
(XRK) Back plate adaptor
(XRH) Calibration report
(XCL) Nonstandard
           calibration
Back

 

1-877-ASHCROFT Credit Cards Accepted by PSI

.

-© FLW, Inc 1997- - S.D.G.